濁度計・色度計(上水・排水用)

Turbidimeter/Color Meter

WA 6000

WA 6000 is an integrating sphere type turbidimeter/color meter based on the “Waterworks law”, “Standard methods for the examination of water” and “Testing methods for industrial water (JIS K0101). A cell of 5mm to 100mm is usable, so you can measure a sample from low to high turbidity range. Thanks to its double-beam measuring method and automatic calibration function, high accuracy and stability are maintained even for long hours of measurement work. This model will be a most suitable instrument to control water quality for a wide range from raw water to clean water.

Features

This is a turbidimeter/color meter based on the “Waterworks law”, “Standard methods for the examination of water” and “Testing methods for industrial water.

Measurement is possible for the turbidity of 0 to 100 NTU (by polystyrene) and 0 to 1,000 NTU (by kaolin), and chromaticity of 90 to 1,000

Highly reliable measurement is available as pre-shipment calibration has been done using turbidity and color standard solutions.

This model has such calculation functions as CV value, standard deviation, correlation coefficient, etc. Multiple working curve data can be also input.

It is an all-in-one type instrument that consists of operation panel, fluorescent character display and built-in printer.

It is possible to connect an auto sampler (option) to this model, by which multiple samples can be automatically measured and labor is saved.

Using data transfer software (option), you can transfer measurement data to your personal computer.

Auto Sampler for WA 6000

WA 2050S

WA 2050S is an auto sampler to be used in combination with WA 6000 for automatic and laborsaving multiple measurements. 50 tall beakers can be set on its turn-table. With one-button operation, calibration, water feeding, measurement, drain and cleaning are automatically repeated. This model has an optional expandability of adding pH measuring function and TOC measuring function, etc.

Features

Automatic and laborsaving multiple measurements are available when this auto sampler is connected with WA 6000.

Approx. 90 sec. high speed measurement is possible for one sample’s color and turbidity measurements (approx. 5 minutes for the pH specification in option).

A job sequence is fully automated for calibration, water feeding, washing by sample water, measurement, drain and cleaning.